Review reviewCritical

CVE-2026-0611

Spacelabs Healthcare Sentinel

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made networ...

CVSS
9.2
EPSS
0.66%
48.3% percentile
CISA KEV
Not listed
Published
2026.06.03
PRIORITY ASSESSMENT

Review review

The CVSS severity warrants an early asset and exposure review.

Known exploitationNot established by KEV
Exploit probability0.66%
Technical severityCVSS 9.2

Vulnerability overview

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made networ...

Affected product and versions

Product
Spacelabs Healthcare Sentinel
Affected versions
>= 10.5.0 < 11.6.0
Fixed versions
No verified fixed-version field is available yet

Recommended response sequence

Confirm exposure before applying a vendor-supported change.

Full remediation guide
  1. 1
    Identify

    Confirm that Spacelabs Healthcare Sentinel and an affected version are present.

  2. 2
    Prioritize

    Combine exploitation signals with asset exposure and business criticality.

  3. 3
    Remediate

    Follow the vendor advisory or supported update path and preserve rollback options.

  4. 4
    Verify

    Recheck the version, service health, access paths, and relevant logs.

Technical data

CVSS vector
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CWE
CWE-306