CVE-2019-25722
Dräger SC 6002XL, SC6802XL, SC 7000 취약점
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient...
- 대응 우선순위
- 점검
- CVSS
- 7.2
- EPSS
- 0.19% 백분위 9.19% · 2026.08.05 기준
- CISA KEV
- 미등록
- 조치 기한
- -
- 공개일
- 2026.06.03