CVE-2019-25720
Dräger SC 6002XL, SC6802XL, SC 7000 취약점
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
- 대응 우선순위
- 점검
- CVSS
- 7.1
- EPSS
- 0.20% 백분위 9.95% · 2026.08.05 기준
- CISA KEV
- 미등록
- 조치 기한
- -
- 공개일